Cleaning and Disinfection Blood Pressure Cuffs

Reusable NIBP Cuffs, tubing and port fittings

 Remove fitting and tubes from cuff,
seal open port with accessory plug. Close the
hook and loop.
Use one or more of the following methods and allow to air
dry:
 Wipe with mild detergent and water solution
(1:9 solution). Rinse.
 Wipe with Enzol® per manufacturer’s
instructions. Rinse.
 Wipe with 0.5% bleach and water solution.
Rinse.
 Wipe with 70% isopropyl alcohol.
 Launder with mild detergent in warm water (60°
C max), normal wash cycle. Cuff is compatible
with 5 wash cycles.

 

NIBP Cuffs Low-Level Disinfection Procedure

Prepare Enzol ® enzymatic detergent according to
manufacturer’s instructions. Apply port-cap to cuff.
Spray detergent solution liberally onto cuff and use a
sterile brush to agitate the detergent solution over entire
cuff surface for five minutes. Rinse continuously with
distilled water for five minutes. To disinfect, first follow
the cleaning steps above, then spray cuff with 10%
bleach solution until saturated, agitate with a sterile brush
over entire cuff surface for five minutes. Rinse continuously
with distilled water for five minutes. Wipe off
excess water with sterile cloth and allow cuff to air dry.

 

Disposable NIBP Cuffs, tubing and port fittings

Use one or more of the following methods and allow to air
dry:
 Wipe with mild detergent and water solution
(1:9 solution). Rinse.
 Wipe with Enzol per manufacturer’s
instructions. Rinse.
  Wipe with 0.5% bleach and water solution.
Rinse.
 Wipe with 70% isopropyl alcohol.

Cleaning and Disinfection Blood Pressure Cuffs

Reusable NIBP Cuffs, tubing and port fittings

 Remove fitting and tubes from cuff,
seal open port with accessory plug. Close the
hook and loop.
Use one or more of the following methods and allow to air
dry:
 Wipe with mild detergent and water solution
(1:9 solution). Rinse.
 Wipe with Enzol® per manufacturer’s
instructions. Rinse.
 Wipe with 0.5% bleach and water solution.
Rinse.
 Wipe with 70% isopropyl alcohol.
 Launder with mild detergent in warm water (60°
C max), normal wash cycle. Cuff is compatible
with 5 wash cycles.

 

NIBP Cuffs Low-Level Disinfection Procedure

Prepare Enzol ® enzymatic detergent according to
manufacturer’s instructions. Apply port-cap to cuff.
Spray detergent solution liberally onto cuff and use a
sterile brush to agitate the detergent solution over entire
cuff surface for five minutes. Rinse continuously with
distilled water for five minutes. To disinfect, first follow
the cleaning steps above, then spray cuff with 10%
bleach solution until saturated, agitate with a sterile brush
over entire cuff surface for five minutes. Rinse continuously
with distilled water for five minutes. Wipe off
excess water with sterile cloth and allow cuff to air dry.

 

Disposable NIBP Cuffs, tubing and port fittings

Use one or more of the following methods and allow to air
dry:
 Wipe with mild detergent and water solution
(1:9 solution). Rinse.
 Wipe with Enzol per manufacturer’s
instructions. Rinse.
  Wipe with 0.5% bleach and water solution.
Rinse.
 Wipe with 70% isopropyl alcohol.

Older Post Newer Post